FED-STD-791D
elements to maintain the test temperature. There shall be one thermocouple provided for each
heating element. Figures 1, 2, and 3 show a satisfactory heating block.
4.4 Temperature controller. A suitable temperature controller capable of maintaining
temperatures between 99 to 538C " 0.5C. A suitable temperature recorder shall also be
provided.
4.5 Flowmeter. A rotameter calibrated to deliver air at a rate of 2.58 " 0.02 g/min
between 15.6 to 29.4C (60 and 85F) (2 L/min at standard temperature and pressure). It shall
be furnished with a needle valve and may be mounted as shown in Figure 7.
4.6 Evaporation tester stand. A suitable stand for supporting the test apparatus is shown
in Figure 5.
5. MATERIALS
5.1 CAUTION - SOME MATERIALS ARE TOXIC AND HAZARDOUS. The
materials references and standards listed in this section must be handled carefully. Federal Test
Method 10000, Material Handling Safety Precautions, is a reference which lists all toxic and
hazardous materials cited in FTMS 791. The synonyms, life hazard, flammability, handling and
storage precautions, emergency treatment and measures, and spill practices of each chemical are
explained.
5.2 1,1,1-Trichloroethane (ASTM D 4126), [2,1,0].
5.3 Tissue, Facial.
6. PROCEDURE
6.1 Greases
6.1.1 Clean the grease-sample cup and hood thoroughly, using warm
1,1,1-trichloroethane (38-54C [100-130F]) and facial tissues. Dry thoroughly in warm air.
Caution. 1,1,1-trichloroethane is toxic. Use it only in a well ventilated area.
NOTE 1:
Avoid contact with skin.
6.1.2 Weigh the clean grease-sample cup and hood to the nearest mg. Remove the hood
and fill the cup with sample, taking care to avoid occlusion of air. Smooth the surface level with
the rim of the cup with a straight edged spatula. Remove with a clean cloth any grease which
may remain on the rim or threads of the cup. Thread the hood tightly onto the cup without
disturbing the smoothed grease surface. Weigh the assembly and record the net weight of the
sample to the nearest mg.
6.1.3 With cover in place, but without the hood and sample cup attached, allow the
evaporation cell to acquire the temperature of the block (controlled to 2C ["1F]), at which the
35