FED-STD-791D
Method 5329.2
September 30, 1986
CORROSION-PROTECTION (HUMIDITY CABINET)
1. SCOPE
1.1 This method is used for determining the effectiveness of corrosion-inhibiting
oils and compounds under conditions of high humidity.
2. SUMMARY
2.1 The test consists of dipping steel test panels in the test sample and then
placing them in a humidity cabinet. After the specified exposure time the panels are
removed and examined for the presence of rust.
3. SAMPLE SIZE
3.1 Approximately 500 mL of the oil to be tested.
4. REFERENCES, STANDARDS, AND APPARATUS
4.1 Standards.
4.1.1 ASTM D 1748 Rust Protection By Metal Preservatives In the Humidity
Cabinet.
4.2 Humidity cabinet conforming to Appendix I of ASTM D 1748 where the
following standard conditions must be maintained.
a. Air temperature:
Inside the cabinet: 49 1C (120 2F).
Outside the cabinet: 24 5C (75 10F).
b. Rate of air to the cabinet: 868 28 L (31 1 ft3) per hour at 25C (77F) and
76 cm of mercury.
c. Water in cabinet:
Level - 20.3 0.6 cm (8 1/4 in)
pH - 5.5. to 7.5
Oil content - clear with no evidence of oil
Chlorides - less than 20 ppm
Sulfates & sulfites - less than 20 ppm.
d. Speed of rotating stage - 0.33 0.03 rpm.
e. Cover - close fitting.
f. Cloth layers in cover - shall not be torn, contaminated, nor contain droplets of
water.