4.5 Vacuum pump, capable of reducing the pressure to at least 1.0 mm of mercury.
4.6 Water bath, capable of maintaining a temperature of 80 2C (175 5F) and
sufficient size to contain the test cell.
4.7 Laboratory safety shield, constructed of either safety glass or wire reinforced plastic.
5. MATERIALS
5.1 CAUTION: SOME MATERIALS ARE TOXIC AND HAZARDOUS. The
materials references and standards listed in this section must be handled carefully. Federal Test
Method 10000, Material Handling Safety Precautions, is a reference which lists all toxic and
hazardous materials cited in FTMS 791. The synonyms, life hazard, flammability, handling and
storage precautions, emergency treatment and measures, and spill practices of each chemical are
explained.
5.2 Cleaning solution glass (concentrated sulfuric acid [3,0,2] saturated with potassium
dichromate, technical grade).
5.3 Detergent for cleaning glassware [u,u,u].
6. PROCEDURE
6.1 Cleaning test cell. Clean the test cell by scrubbing with a suitable detergent and rinse
with distilled water. Prior to the introduction of the sample into the test cell, the cell shall be
rinsed in hot potassium dichromate-sulfuric acid cleaning solution, thoroughly rinsed in distilled
water, and dried in an oven at 107 2.5C (225 5F) for 1 h.
6.2 Preheat and adjust heater temperature to 260 1C (500 2F).
6.3 Perform duplicate test procedures of 6.4 through 6.13.
6.4 Filter the sample through a nominal 10-m filter paper to remove particle
contaminants.
6.5 Introduce 20 mL of the filtered sample below the constricted area of the test cell with
the aid of the hypodermic syringe.
NOTE: This syringe is capable of delivering 20.0 0.5 mL of sample which is within the
accuracy required by this test procedure.
6.6 Connect the filled test cell to the vacuum pump and degas the cell and its contents,
while shaking gently for at least 5 min or until foam no longer persists.
6.7 Place the test cell, while still connected to the operating vacuum pump, in a water
bath maintained at 80 2C (175 5F) for 5 min. Remove the cell from the bath, shake
73